52L Laboratory Refrigerator | Model L 55
Laboratory Refrigerators, such as the B Medical Systems 52L Laboratory Refrigerator | Model L 55, are for the safe storage of samples, specimens, cultures and other laboratory preparations at cold temperatures between +4°C and +15°C with multiple controls and warning safety features developed as a result of 40 years of expertise in refrigeration technology.
Integrated multi-function monitoring electronics
Integrated B Medical Systems multi-function electronics, at easy access level in the door; offers an alarm system, setpoint security, battery backup, full display, and many other control features; compatible with °B Connected solution for remote data monitoring and recording.
Lifetime warranty corrosion-free cabinet
• One-piece material cabinet with a lifetime warranty for zero corrosion, highest quality polypropylene for better insulation and no sharp edges to provide easier maintenance and cleaning
• Smart ergonomy – Easier and safer handling because of heavier components placed at the bottom
High storage capacity
High storage per square meter in terms of net volume, with a very versatile inner volume utility due to the multiple user-friendly rails allowing modularity.
Specifications
- Gross / Net volume (l) 52/45
- Set temperature (preset) +5°C
- Set temperature (setting range) can be adjusted in steps of 0.1°C +4°C to +15°C
- Temperature cold / warm alarm limit +3°C / +7°C
- Hold over time (+5°C to +10°C) 1.3 h
- Climate class (ambient temperature range) SN (+10°C to +32°C)
- Defrosting technique Natural
- Refrigerant type R600a
- External dimensions H x W x D (mm) 670 x 495 x 575
- Inner dimensions H x W x D (mm) 470 x 375 x 350
- Net weight with standard equipment (kg) 38
-
Supply voltage (V) 220-240
-
Frequency (Hz) 50/60
-
Power (W) 100
-
Energy consumption (kWh/24h) 0.38 / 0.45
-
Heat emission (Kcal/h) 9
-
Compressor running time (%) 25
-
Noise level (dB(A)) (at 1m height & 1m distance) 35/36
-
Supply voltage (V) 220-240
-
Frequency (Hz) 60
-
Power (W) 100
-
Energy consumption (kWh/24h) 0.38
-
Heat emission (Kcal/h) 8
-
Compressor running time (%) 20
-
Noise level (dB(A)) (at 1m height & 1m distance) 37